y The LEDs next to the display indicate the results:
yellow LED: sample too heavy
green LED: sample in tolerance range
red LED: sample too light
p t Print the results.
Note: If automatic printout of results is enabled, you do not need to press the
p key. The results are printed automatically.
Printout configuration: see page 96
Setp + 1.250 kg Target
Min + 1.240 kg Minimum
Max + 1.280 kg Maximum
G# + 1.256 kg Gross weight
T + 0.000 kg Tare weight
N + 1.256 kg Net weight
Lim + 0.48 % Percentage of deviation from target*
W.Diff+ 0.006 kg Absolute deviation from target
-------------------
* When displayed in relation to target value:
If the weight is lighter than the target, the display shows: LL
If the weight is heavier than the target, the display shows: HH
Example 2: Checkweighing samples with a target weight of 1250 g and a tolerance range
from -10 g to +30 g.
The tolerance values should be entered as a relative deviation from the target value.
Configuration: The “Checkweighing" application is selected using the setting INPUT
/ TAR.TOL, a printout has been set up (see “Configuration").
O t Start target value and tolerance entry using the O key.
y The target value symbol flashes at the top of the display.
t Place a sample with the target weight (in this example, 1250 g) on the platform.
O t Save the target value.
y The minimum symbol flashes at the top of the display.
Operating Instructions Combics Indicators 77
Operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CAISL1 and is the answer not in the manual?

Sartorius CAISL1 Specifications

General IconGeneral
BrandSartorius
ModelCAISL1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals