EasyManuals Logo

Sartorius CIS1 User Manual

Sartorius CIS1
121 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
81
Configuring the A/D Converter (ADC)
Open the menu access switch.
§ Remove the cap that covers the menu access switch on the left-hand side of the back
of the indicator housing.
§ Move the switch to the right (towards the interface connectors; “open" position)
Activate the service mode (see page 77)
Select the weighing instrument to be configured (in this example, WP-1)
Under menu item 9.1, select “standard" (9.1.3) or “trade" (9.1.4) configuration
Open menu item 11 to configure the A/D converter
Following configuration, select menu item 11.10.1 to save your settings.
The Combics is automatically re-started.
The A/D converter, in conjunction with the weighing instrument, can now be used like
any standard weighing instrument.
Close the menu access switch
Following A/D converter configuration, the weighing instrument must be calibrated/
adjusted (perform calibration/adjustment and linearization; see page 20, “Calibration and
Adjustment," and page 89, “Calibration/Adjustment without Weights" for details)