Error Code  Explanation 
Err 220  ROM checksum error; the data in the internal ROM 
of the balance processor (AOC) are  incorrect. 
Err 230  RAM read/write error; the data in the internal RAM 
of the balance processor AOC is incorrect, or not 
possible. 
Err 237  EEPROM checksum error in the linearity range; the 
balance has not yet been linearized, or the data in 
the internal EEPROM of the balance processor AOC 
are incorrect. 
Err 239  EEPROM checksum error in the linearity weight
  range; the stored factor for the linearity weight 
is not within the permissible tolerance range or is 
incorrect. 
Err 241  EEPROM checksum error in the permanent area; 
the data for the balance operating menu in the
 internal EEPROM of the balance processor AOC are 
incorrect. 
Err 243  EEPROM checksum error in the menu range;
  the inalterable data in the internal EEPROM    
  of the balance processor AOC are incorrect.
Err 245  EEPROM checksum error in the calibration/ 
  adjustment range (zero point); the balance has not 
yet been calibrated or the data in the internal 
  EEPROM of the balance processor AOC are incorrect. 
Err 247  EEPROM checksum error in the adjustment/ 
  calibration range (span); the balance has not yet 
been calibrated or the data in the internal EEPROM 
of the balance processor AOC are incorrect. 
Err 249  EEPROM checksum error in the calibration weight 
range; the factor stored for the internal weight cali-
bration exceeds the tolerance range, or is incorrect.
 
L  The weight on the pan is below the weighing capa-
city; the pan is not on the balance, the balance was 
incorrectly adjusted or the preload was incorrectly 
adjusted. 
H  The weight on the pan exceeds the weighing capa-
city; the weight is too heavy; the balance was 
incorrectly adjusted; or the preload was incorrectly 
adjusted.
 
Remedial Measures
Follow the procedure in “Exchanging the Main PCB.”
Follow the procedure in “Exchanging the Main PCB.”
Overwrite the incorrect data record using the MC1 Server 4.9. 
Then proceed as directed in „Adjusting the Linearity” and 
“External Calibration.” 
Follow the procedure for “Overwriting the Internal Linearity
Weight.”
Overwrite the incorrect data record using the
MC1 Server 4.9.  If the error remains, then proceed with 
“Exchanging the Main PCB.”
With the menu access switch unlocked, access the menu 
(display will fl ash). Exit the menu and store (press the tare key
for more than 2 sec.) => The menu is now reset. Checksums
are corrected.
Perform “External Calibration/Adjustment.”
Perform “External Calibration/Adjustment.”
Follow the procedure for “Overwriting the Internal 
Calibration Weight.”
Make sure there was no operating error.  If the error remains,
perform “External Calibration/Adjustment” or follow the 
procedure for “Adjusting the Preload.”
Make sure there was no operating error.  If the error remains,
perform “External Calibration/Adjustment” or follow the 
procedure for “Adjusting the Preload.”
 | 41