8 |
Auxiliary Service Tools, Equipment and Weights
Caution: After removing the seal (warranty label) and fi nishing service work, reaffi x your 
control seal with the Sartorius logo!
Please make sure that you use the proper tools and equipment and that the 
balance/scale is set up on a solid, level surface and in a clean work area that is free 
of vibrations and drafts.
PSION CAS Version 4.9 and later
Service software CAS for PC Version 1.44 or later   6740-33
and
    RS232 connecting cable (25-Pin)    7357312
or
    RS232 connecting cable (9-Pin)    7357314
Tool Kit  Angled fi le    6740-80
In addition to your standard tools and equipment, you will need the following sets of 
weights when working on Sartorius Competence balances/scales:
For Model      Accuracy    Weight    Catalog No.:
CPxxxx-CPAxxxx    class (OIML)  in grams
CP64        E2   1x50   YCW4528
CP124S, GC803 S/P      E2     1x100     YCW5128
CP323P, CP153        F1     1x100     YCW5138
CP225D, CP224S         E2     1x200     YCW5228
CP324S, GC1603P       E2     1x200+    YCW5228 +
      1x100   YCW5128
CP423S, CP323S, 
GC 2502          F1     1x200     YCW5238
CP622           F2     1x500     YCW5548
CP3202P, CP2202S      F1     1x1000    YCW6138
CP4202S, CP3202S,
CP4201, CP2201,
GP3202          F1     1x2000    YCW6238
CP8201, GP8201,
CP6201          F1     1x5000    YCW6538
CP34001S, CP34001P,
CP16001S, CP12001S,
CP34000         F1     1x10000   YCW7138
Accompanying Literature
Operating Instructions Sartorius Competence/CP/CPA Series/Sartorius Gemplus Publica-
tion No: WCP6006-e05073
Auxiliary Service Tools, Equipment and Weights