32 |
- Make sure that the balance/scale operating menu is set to the code „1 9 3 -
internal calibration/adjustment accessible „ (see Page 13-14 „ Operating Menu
Settings „).
- Slide the access lock switch to the left. It is accessible via an opening on the
back of the balance/scale.
- If you have done so already, perform external calibration/adjustment (see Page
31).
- Turn off the balance/scale by pressing e.
- Turn the balance/scale back on by pressing e and while simultaneously
holding down the q key until the readout shown on the left appears.
- Press the t key to tare the balance/scale; then press q to activate the
„overwrite internal calibration weight“ function.
- The built-in, motorized internal calibration weight is applied, then removed
automatically; the weight value is overwritten; and the balance/scale returns to
the normal weighing mode.
- Press the q key to check the external span adjustment once again.
Internal Span Adjustment
- Make sure that the menu code „1 9 3 - internal calibration/adjustment acces-
sible“ is set (see Page 13 -14 Operating Menu Settings).
- Press the t key to tare the balance/scale; then press q to activate the
calibration/adjustment function..
- The built-in, motorized internal calibration weight is applied, then removed
automatically; the weight value is overwritten; and the balance/scale returns to
the normal weighing mode.
- Slide the access lock switch back to the right. Replace the capped plug to close
the opening.