ISO/GLP-compliant Printout/Record
Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series. These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance/scale manufacturer
– Balance/scale model
– Balance/scale serial number
– Software version number
– Identification number of the current
sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
The record is output to a Sartorius
data printer or a computer.
52
Settings
● Set the following menu codes
(see “Configuring the Balance/Scale”):
– ISO/GLP-compliant record
after calibration/adjustment only:
menu code 8 10 2; or
ISO/GLP-compliant record always on:
menu code 8 10 3
– Line format for printout: With data ID codes
– 22 characters: menu code 7 2 2
!No ISO/GLP-compliant record is output if
any of the following settings are configured:
menu codes 6 1 4, 6 1 5, 6 1 6 (automatic
printout) and 7 2 1
Function Keys
Press p to output header and first
measured value.
> Header is output the first time p
is pressed
To output header and reference data
automatically with an application program
active: Press v
End an application:
1) Output GLP footer: Press c
2) End application program: Press c