52
Error Codes
Display Cause Solution
E 22 Weight is too light or there is no Increase the weight on the balance/
sample on the balance/scale scale
E 30 „Print“ key pressed while balance/ Activate SBI mode (e.g., using the
scale is in BPI mode „Close“ function in the service software)
E 50 Overload  or  underload  in temperature Defect in TK sensor or main PCB
compensation converter (replace)
E 53 Temperature  compensation  converter Defect in TK sensor or main PCB
not functioning.  No measured value (replace)
received in processor from TK switch
E 54 A/D converter output is below the Device is underloaded; no weighing
minimum level pan/ load plate on balance/scale;
mechanical defect; defective main PCB
E  55 A/D converter output is above the Device is overloaded; mechanical
maximum level defect; defective main PCB