Talent and Gold Series
Model TE313S, TE153S, TE3102S, GE2102
TE313S-DS TE153S-DS GE3102
Weighing capacity g 310 150 3,100 2,100
Readability g 0.001 0.001 0.01 0.01
Tare range (subtractive) g 310 150 3,100 2,100
Repeatability <±g 0.001 0.0015 0.01 0.015
Linearity <±g 0.002 0.003 0.02 0.04
Operating temperature range +10…+30 °C (50° to 86°F)
Allowable ambient
operating temperature +5… +40°C (41°F to 104°F)
Sensitivity drift <±
within +10…+30 °C /K 4·10
–6
4·10
–6
3·10
–6
4·10
–6
Response time (average) s 3 2.5 2.5 2.5
Adaptation to
ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends
on the filter level selected) s 0.2–0.8
External calibration weight g 200 (E2) 100 (F1) 2000 (E2) 2000 (F1)
(of at least accuracy class) lb* 0.4 0.2 4 –
Net weight, approx. kg 2.2/3.2 1.7/2.7 2.2 1.7
Pan size mm 100 d 100 d 174+143 174+143
Dimensions (B+T+H) mm 200+270+120/ 200+270+70
TE…-DS:
200+270+299
AC power source/
power requirements AC adapter 230 V or 115 V, +15% to –20%
Frequency 48–60 Hz
AC power source/
direct current V 10–20
Power consumption
(average) W 0.75
Hours of operation with
the YRB08Z rechargeable
battery pack h 20 25 20 20
* = only for TE models
26