– Internal calibration/adjustment:
Internal cal./
adjustment
– Internal linearization:
Internal lin.
– Repeatability test
reproTEST
– External calibration/adjustment with
the preset calibration weight:
Ext. cal./adj.;
factory-def. wt.
51
Selecting the Calibration/Adjustment
Parameter
The setting Selection mode
must be selected in the Setup menu
(factory setting). After pressing the
Cal soft key, you can choose
from among the following settings by
pressing the
Select soft key:
– External calibration/adjustment with a
calibration weight determined by the
user:
Ext. cal./adj.;
user-defined wt.
● Start the desired routine:
Press the
Start soft key
In the selection mode: Perform external calibration followed by automatic adjustment with the factory-set weight
Configuration:
factory settings
Step Press key(s) (or follow instructions) Display/Output
1. Select calibration
Cal soft key
2. Select external calibration/adjust-
Select soft key 3 x
ment with factory-defined weight
3. Start external calibration/
Start soft key
adjustment
4. Place the weight on the Place std. weight on balance
balance (e.g., 200.00000 g)
Minus sign –: Weight too low
Plus sign + Weight too high
no plus/minus sign: Weight o.k
This is displayed after calibration,
for approx. 10 seconds:
(on verified balances, the
difference between the displayed
weight and the conventional
mass value is displayed)
After adjustment, the following is displayed
5. Unload the balance
(ISO/GLP printout: see page 131)