EasyManua.ls Logo

Sartorius LA230S - DKD Uncertainty of Measurement

Sartorius LA230S
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
109
Example
Perform a weighing procedure with the “DKD uncertainty of measurement” application
Settings:
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Print: PA (process accuracy)
Setup: Application parameters: Extra function (F4): DKD uncertainty of measurement: Display: Input: 3.00000 (factory setting)
Step Press key(s) (or follow instructions) Display/Output
1. Turn on the balance, if not on, and e
configure the settings as indicated above
2. Place a container for a sample on the w
balance and tare
3. Measure weight of sample Add sample to container
4. Print weight r U 0.08 g
PA 0.024 %
N +2000.2 g
5. Weigh next sample (if any)

Related product manuals