EasyManua.ls Logo

Sartorius LA230S - Specifications of Verified Models (LA1200 S-0 CE to LA2200 S-0 CE)

Sartorius LA230S
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
138
Model LA1200S-0CE LA620S-0CE LA220S-0CE LA620P-0CE
Type isoTEST in conjunction with BD BF
Accuracy class* kKKK
Scale interval, d* g 0.001 0.001 0.001 0.001/0.002/0.005
Maximum weighing capacity, Max* g 1,200 620 220 120/240/620
Verification scale interval, e* g 0.01 0.01 0.01 0.01
Minimum capacity, Min* g 0.1 0.02 0.02 0.02
Tare range (subtractive) 100% of the max. weighing capacity
Application range according to CD* g 0.1 – 1,200 0.02 – 620 0.02 – 220 0.02 – 620
Response time (average) s 1.5
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams, kilograms
External calibration weight value g 1,000 (E2)
(of at least accuracy class...)
Pan size mm d 130
Dimensions (W + D + H) mm 261 + 381 + 147
Net weight, approx. kg 8.3 6.9 6.9 6.9
Dust and water protection rating
according to EN 60529
2
) IP54
Model LA8200S-0CE LA8200P-0CE LA6200S-0CE LA4200S-0CE LA2200S-0CE
Type isoTEST in conjunction with BD BF
Accuracy class* KKKKK
Scale interval, d* g 0.01 0.01/0.02/0.05 0.01 0.01 0.01
Maximum weighing capacity, Max* g 8,200 2,000/4,000/8,200 6,200 4,200 2,200
Verification scale interval, e* g 0.1 0.1 0.1 0.1 0.1
Minimum capacity, Min* g 0.5 0.5 0.5 0.5 0.5
Tare range (subtractive) 100% of the max. weighing capacity
Application range according to CD* g 0.5 – 8,200 0.5 – 8,200 0.5 – 6,200 0.5 – 4,200 0.5 – 2,200
Response time (average) s 2 2 1.5 1.5 1.5
Allowable operating temperature range 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F) with the isoCAL function
1
)
Selectable weight units Grams, kilograms
Pan size mm 218 + 200
Dimensions (W + D + H) mm 261 + 381 + 86
Net weight. approx. kg 6.5
Dust and water protection rating
according to EN 60529
2
) IP54
1
) = With the isoCAL function deactivated, the verified balance can only be used within the limited temperature range (can only be modified
by the Sartorius Service Center): For balances of accuracy class k: +15°C to +25°C (+59°F to +77°F)
For balances of accuracy class K: +10°C to +30°C (+50°F to +86°F)
2
) = Specially protected dust-tight and washdown-resistant AC adapter; see the section on “Accessories.”
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area

Related product manuals