EasyManua.ls Logo

Sartorius LA620S - Display and Control Unit Output; Metrological Data Line Display

Sartorius LA620S
149 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
113
Max4200 g
Min 0.5 g
e= 0.1g
d=0.01g
J
F H
S
There are 3 options for data output:
– Output to the display and control unit
– Output to a printer (generate a printout)
– Output to a peripheral device (e.g., computer) via the interface port
Output to the Display and Control Unit
The display is divided into 9 sections. Information about the balance, the application
being used and the sample weighed is output in the following sections:
– Line for metrological data
– Bar graph
– Plus/minus sign, stability symbol display
– Measured value line
– Weight unit display
– Data in tare memory; calculated value
– Application symbol display
– Text line
– Soft key labels
Line for Metrological Data (on balances verified for legal metrology)
This line shows:
– Maximum balance capacity (e.g., 4,200 g)
– Minimum balance capacity; the weight must not go below this limit when the balance
is used in legal metrology
– Verification interval of the balance; irrelevant if the balance is not used in legal metrology
(e.g., 0.1 g)
– Readability: Indicates the actual scale interval (display increment of the balance)
(e.g., 0.01 g)
Bar Graph (overview display)
In the bar graph, weighing results are displayed either
– as a percentage of the maximum balance capacity, or
– in relation to a target value, with tolerance limits indicated.
You can turn off (blank) the bar graph display (Setup: Device: Display: Digit size:
13 mm + text display or 13 mm)
Plus/Minus Sign, Stability Symbol
This section shows:
– “Busy” symbol
– Plus or minus sign
– Zero symbol (indicating the scale has been zeroed)
Data Output Functions