Setting the Drying Parameters
Purpose
Adapt the moisture analyzer to the special requirements of products.
Parameters can be configured individually for every program.
Drying Parameters (Overview)
ο Factory setting
User-defined setting
Program Program name None
memory
(Prog.) Heating program ο Standard drying 105°C 30 to 200°C
Quick drying 105°C 30 to 200°C
Gentle drying* Temperature: 105°C 30 to 200°C,
Time: 3.0 min. 1st phase 1.0 to 20 minutes
Phase drying* Temperature 1: 80°C 30 to 200°C,
Time 1: 5.0 min. 0.0 to 99.9 minutes
Temperature 2: 105°C 2nd phase: 30 to 200°C,
Time 2: 5.0 min. 0.0 to 99.9 minutes
Temperature 3: 120°C 3rd phase: 30 to 200
High-temperature drying 105°C 30 to 230°C
Standby temperature ο Off
On Temperature: 40°C 30 to 100°C
Initial weight ο Off
Minimum and Max. init.wt.: 5 g MA100: 1 mg to 90% of max. capacity
maximum initial weight MA50: 10 mg to 90% of max. capacity
Min. init.wt.: 1 g MA100: 1 mg to 90% of max. capacity
MA50: 10 mg to 90% of max. capacity
Target weight Target weight: 5 g MA100: 2 mg to 90% of max. capacity
Tolerance in % MA50: 20 mg to 90% of max. capacity
Tolerance: 10% 1 to 50%
Start analysis ο With stability + Time delay: 2 sec. MA100: 0 to 99 seconds
auto close MA50: 2 seconds only
With stability, Time delay: 2 sec. MA100: 0 to 99 seconds
manual close MA50: 2 seconds only
Without stability, Time delay: 2 sec. MA100: 0 to 99 seconds
auto close MA50: 2 seconds only
Without stability, Time delay: 2 sec. MA100: 0 to 99 seconds
manual close MA50: 2 seconds only
Fully automatic, No
With stability time delay
Fully automatic, No
Without stability time delay
End of analysis ο Fully automatic
asap: determ.-
semi-auto
parameters
Semi-auto: absolute Loss: 10 mg 1 to 50 mg
weight loss Time interval: 60 sec. 5 to 300 seconds
Semi-auto: weight Loss: 1.0 % 0.1 to 50 %
loss in percent Time interval: 60 sec. 5 to 300 seconds
Time Time: 15.0 min. 0.1 to 999.9 minutes
Manual
* = for MA100 only
30
Factory setting
Factory setting