EasyManua.ls Logo

Sartorius MA50-H - Page 50

Sartorius MA50-H
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step Key (or instruction) Display/Printout
6. Prepare the sample: not necessary
for dispersion adhesive
7. Open the sample chamber k key
Position a new sample pan
8. Tare the sample pan
Tare soft key
9. Apply approx. 1.5 g of
dispersion adhesive uniformly
to the sample pan
Close the sample chamber k key
10. Start the drying program
Start soft key
After 2 seconds’ delay --------------------
the header for moisture 14.07.1999 17:10
analysis is printed
Mod. MA100C
Ser. no. 90805355
Ver. no. 01-38-07
ID
--------------------
Prg 5 DISP ADHESIVE
Heating PHASES
Temp.1 140
C
Time1 5.0 min
Temp.2 140
C
Time2 0.0 min
Fin. temp. 140
C
Stdby temp. OFF
Start W/STABIL.
End AUTOMATIC
IniWt + 1.502 g
--------------------
The current moisture loss
displayed afterwards
49
Info ID Stat. Mode Prog.
P5 140/140/140 C Auto.
WEIGH-IN: Load sample pan
+ 0000 g
Info ID Stat. Mode Prog.
P5 140/140/140 C Auto.
WEIGH-IN: Start analysis
+ 1502 g
Start
ID Stat. Para. Prog.
P5 140/140/140 C Auto.
ANALY.BEGIN: Start delay 2sec
+ 1502 g
Abbr.
S
Info ID Stat. Mode Prog.
P5 140/140/140 C Auto.
ANALYSIS: 27 C 0.3min ----mg/24s
+ 006 %L
Cancel

Related product manuals