Interface Description
Purpose
The moisture analyzer has an interface
port for connection to an external
printer or computer (or other peripheral
device).
External Printer
You can use an external printer to gen-
erate printouts and document settings.
Computer
Analyses and calculated values can be
transmitted to a computer for further
evaluation and recording.
You can use a computer to monitor and
remote-control the weighing system
functions.
! Warning When Using Pre-wired
RS-232 Connecting Cables!
RS-232 cables purchased from other
manufacturers often have incorrect pin
assignments for use with Sartorius
weighing systems! Be sure to check the
pin assignments against the chart
below before connecting the cable and
disconnect any lines marked “Internally
Connected” (e.g., pin 6). Failure to
do so may damage or even completely
ruin your weighing system and/or
peripheral device.
61
Available Features
Type of interface: Serial interface
Operating mode: Full duplex
Standard: RS-232
Transmission rates: 150; 300; 600; 1,200; 2,400; 4,800; 9,600; 19,200 baud
Number of data bits 7, 8 bit
Parity: Space, odd, even
Number of stop bits: 1 or 2 stop bits
Handshake mode: Software, hardware 1 character
Operating mode: SBI, XBPI
1
), YDP01IS
2
), YDP01IS-Label
2
), Sartonet
Network address
4)
: 0, 1, 2, ..., 30, 31
Data output format of the 22 characters + CR LF
weighing system:
1
) XBPI operating mode: 9,600 baud, 8 bits, odd parity, 1 stop bit
2
) YDP01IS and YDP01IS-Label operating mode: 9,600 baud, 8 bits,
odd parity, 1 start bit, 1 stop bit, software handshake
3
) Network address is only valid in the XBPI and Sartonet modes
Parameter Factory Settings for the Interface Port:
Transmission rate: 1,200 baud
Number of data bits 7 bits
Parity: Odd
Stopbits: 1 stop bit
Handshake: Hardware; 1 character
Operating mode: SBI
Preparation
See pages 66 and 67 for the pin assignment chart and cabling diagram.