Do you have a question about the Sartorius Master Pro LP620P and is the answer not in the manual?
| Capacity | 620 g |
|---|---|
| Readability | 0.001 g |
| Calibration | External |
| Linearity | ±0.002 g |
| Display | LCD |
| Power Supply | AC Adapter |
| Repeatability | 0.001 g |
| Stabilization Time (typical) | 2 s |
| Interface | RS-232 |
| Operating Temperature | +10°C to +30°C |
Prerequisites for working on Sartorius Masterpro Balances, including experience and tools.
Explanation of the balance's operational keys like ON/OFF, Tare, Print, Setup, etc.
Accessing and navigating the balance's setup menu.
Entering calibration weight values and decimal places for adjustment.
Inputting values for 1st to 4th linearization weights in service mode.
User/Service access for entering data, weights, and passwords.
Location and function of the access switch for service mode.
Procedure to activate BPI mode for calibration/PCB replacement.
List of adjustment options available in service mode.
Procedure for entering specific weight values for linearization.
Starting external linearization using predefined weights.
Starting external linearization with user-defined weights.
Methods for performing external span adjustment.
Inputting calibration weights for span adjustment when needed.
Procedure to replace the internal calibration weight data.
Defining and using custom weights for calibration.
Testing the reproducibility of the built-in calibration weight.
Procedure to detach the display and control unit.
Steps to open the weighing cell for small single-casting block systems.
Steps to open the weighing cell for large single-casting block systems.
Instructions for replacing the internal calibration weight.
Methods to check preload using service software or a digital voltmeter.
Procedure to adjust preload using a digital voltmeter.
Procedure to check span accuracy using certified calibrated weights.
Procedures for checking linearity based on balance resolution.
Procedure for replacing the PCB in the display and control unit.
Steps for replacing the balance PCB, including data transfer.