Describes three password levels: user, dealer, and service technician access.
Diagram illustrating the flow between User and Service modes for SETUP, INFO, MENU, and INPUT functions.
Lists adjustment options: internal/external adjustment, linearization, reproducibility, and display resolution.
Procedure to manually input linearization weights using the INPUT mode.
Procedure to manually input linearization weights using the INPUT mode.
Procedure to input calibration weight values for span adjustment when using non-standard weights.
Describes key functions like On/Off, Clear, Tare, Print, Toggle, Setup, and Calibration keys.
Factors to check and adjust: preload, reproducibility, off-center load, span, and linearity.
When to check preload and two methods: using service software or a digital voltmeter.
Checking off-center load by placing weights at different positions and comparing values.
Performing a rough 3-point adjustment for off-center load on brass system balances.
Adjusting off-center load on large single-casting block systems by filing the load cell upper guide.
Checking span accuracy using certified calibrated or verified weights.
Procedures for checking linearity based on balance resolution (<1M digits or >1M digits).
Checking linearity for balances with less than 1,000,000 digits using specific test points.
Steps to replace the PCB in the display and control unit, including data transfer.