119
Purpose
This extra function enables you to load
weight values and calculated results
directly from Application 1 (such as
counting, weighing in percent) or
Application 2 (checkweighing, time-
controlled functions) into Application 3
(totalizing, formulation, statistics).
Features
– You can assign this function to the
fourth or fifth soft key (from the right);
i.e, F4 or F5; the soft key label for this
function is
M+
– An Application 3 program (totalizing,
formulation, statistics) must be running
so you can display and print the result
Factory Settings of the Parameters
No user-definable parameters
Saving Values Manually in M+
Preparation
G Turn on the balance: press e
> The Sartorius logo is displayed
G Select Extra function (F4) or Extra function (F5) in the Setup menu: press o
G Select
Application parameters: press the q soft key 2 x,
then the
O soft key
G Select
Extra function (F4) or Extra function (F5):
press the
q soft key 3 x (or 4 x), then the O soft key once
G Select
Man. store in app.3 memory (M+)
G Confirm Man. store in app.3 memory (M+)
See also “Application Parameters (Overview)” in the chapter entitled
“Configuring the Balance”
G Save settings and exit the Setup menu: press the
oo soft key