60
Internal Calibration/Adjustment
First set either 
Internal
cal./adjustment or
Selection mode (factory 
setting) in the Setup menu
(Balance/scale functions:
Calibration/adjustment: CAL key 
function).
Inside the balance housing are built-in
motorized calibration/adjustment
weights.
The internal calibration/adjustment
sequence is as follows:
G Unload the balance
G Select the calibration function: press 
the 
Cal soft key and then the
Start soft key
> The internal calibration weight 
is applied automatically
> The balance is calibrated
> If the setting 
Calibrate,
then auto adjust (factory
setting) is selected in the Setup menu,
the balance is now automatically 
adjusted
> If the setting 
Calibrate,
then manual adjust 
is selected in the Setup menu, 
the internal calibration routine is now
ended without adjusting the balance
(otherwise, see “Calibration and 
Adjustment Sequence,” on next page)
> The internal calibration weight 
is removed
> ISO/GLP printout: 
see page 139
Internal Linearization
In the Setup menu (Balance/scale func-
tions: Calibration/adjustment: CAL key
function), you need to have set
Internal linearization
or select this using the Selection
mode (factory setting).
Inside the balance housing are built-in
motorized calibration/adjustment
weights for linearization.
The internal linearization sequence 
is as follows:
G Unload the balance
G Select the linearization function: press
the 
Cal soft key, the Select soft
key, and then the 
Start soft key
> The internal linearization weights are
applied automatically.
> The balance is linearized
> The internal linearization weights are
removed
> The balance is adjusted automatically
following internal linearization
> ISO/GLP printout: see page 139