Example 3:
Enter or change values for verifiable configuration in single range
mode in the unit set under 1.7.x.
Select menu item ADC CON
) Confirm menu item ADC CON
(if necessary press and select the menu item VERIF.
)k p)
Verifiable configuration
) Confirm menu item VERIF.
Accuracy class
(if necessary press ) ) () To confirm accuracy class 3/4
Range selection
) Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item SINGLE
Multi-interval scale
( Confirm menu item SINGLE
Weights
) Confirm menu item WEIGHTS
Verifiable scale interval
) Confirm menu item E
( ) k p Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly)
until display E appears
k Select menu item MAX.
Enter values for the following in
the same way:
Maximum capacity (e.g. 30 kg)
Continue as shown in example 1
after entering the maximum
capacity
max.
E
Weights
SINGle
Ranges
Class
Verif.
ADC-CON
Example 4:
Enter or change values for verifiable configuration in a multi-inter-
val scale in the unit set under 1.7.x.
(the same applies for multiple range mode).
Select menu item ADC CON
) Confirm menu item ADC CON
(if necessary press
)k p) and select the menu item VERIF.
Verifiable configuration
) Confirm menu item VERIF.
Accuracy class
(if necessary press ) ) () To confirm accuracy class 3/4
Range selection
) Confirm menu item RANGES
(if necessary press k repeatedly) Select menu item MULT. INT
Multi-interval scale
( Confirm menu item MULT. INT
Weights
) Confirm menu item WEIGHTS
Verifiable scale interval
) Confirm menu item E
( ) k p Enter a value (e.g. 0.002 kg)
(if necessary press ) repeatedly)
until display E appears
k Select menu item RANGE 1
Enter values for the following in
the same way:
Range limit 1 (e.g. 6 kg)
Range limit 2 (e.g. 15 kg)
Maximum capacity (e.g. 30 kg)
Continue as shown in example 1
after entering the maximum
capacity
Range 1
E
Weights
Mult.Int
Mult.Int
Class
Verif.
ADC-CON
16

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Midrics 1 and is the answer not in the manual?

Sartorius Midrics 1 Specifications

General IconGeneral
BrandSartorius
ModelMidrics 1
CategoryAccessories
LanguageEnglish

Related product manuals