70 Cubis MSE Operating Instructions
Data Interfaces
Data Output Format with 22 Characters
Here, the data output format with 16 characters is preceded by an ID code
with a 6 characters� These six characters identify the subsequent value�
Normal Operation
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
K K K K K K + A A A A A A A A A * E E E CR LF
* * * * * * � � � � � � � � � * * * *
* * * * * * * * * *
K: ID code character
1
) S: Unit symbol character
1
)
*: Space See Section “Mass Unit Conversion”
A: Displayed characters CR: Carriage return
LF: Line feed
Example:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
N + 1 2 3 5 6 * g * * CR LF
N + 1 2 3 5 [ 6 ]
1
) g * * CR LF
1
) Identification of non-verified digits
Non-verified digits where „e=d“ are automatically marked on the printout by
square brackets when you select the following setting: Configuration:
Communication mode: Printer (code 1� 5� 6� 2)�
SBI setting:
In the SBI setting (code 1� 5� 6� 1), the unverified values are not automatically
identified� Please take appropriate measures or adjust the settings on the
peripheral device�
Special Outputs
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * * * * * * * * * * * * CR LF
H i g h
L o w
C a l E x t
*: Space High: Overload
Cal� Ext�: Calibration, external Low: Underload
Error message
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S t a t * * * * * E R R * # # # * * * * CR LF
S t a t * * * * * A P P � E R R
2
)
* * * * CR LF
S t a t * * * * * D I S � E R R
2
)
* * * * CR LF
S t a t * * * * * P R T � E R R
2
)
* * * * CR LF
*: Space # # #: Error code number
2
) For cause and solution, please refer to „Error Messages“