Application Programs
60  Cubis MSE Operating Instructions
Mass Unit Conversion
Purpose
This application is used to change the weight value displayed from the basic weight unit to any of 4 application weight units 
(see table on next page)�
Features
–  Set the basic unit and display accuracy in the Setup menu: See “Configuration�” 
–  Set the application weight units and display accuracies for all 4 units in the Application menu� 
–  The units selected are saved to protected memory� 
–  The basic unit is always active once the balance is turned on�
Example: Change display from grams [g] (basic unit) to pounds [lb] and then to Troy ounces [ozt] reduced by 1 digit�
Configuration: Applic. - unit 
Step    Press key  Display/Printout
Preparation:
1�  Start selection of an application weight unit  S  unit 1
2�  Confirm unit 1  V  NO °
3�  Select an application unit (in this example: pounds) 
  Units and display accuracy:  Press repeatedly 
  See table on the next page  S  pounds 
4�  Confirm pounds weight unit   V  pounds 
 
°
5�  Select next application unit,  F  unit 2 
  Units and display accuracy: S, V  NO 
 
SDgr
 
  See table on the next page  Repeatedly press S, V  troy.oz. 
6�  Confirm troy ounces weight unit  V troy.oz. °
7�  Reduce display unit by 1  Repeatedly press S, V   minus 1 SDgr
8�   You can select an additional 4 units, if desired  
(otherwise, confirm “NO” using V)
9�  Save selection  Repeatedly press F   0.00 g
  Operation:
10�  Place the sample on the scale    +   100.00 g
 
11�  Change weight unit  Press repeatedly  + 0.22046 lb 
  V +    3.527 ozt