Connect the ex-link converter to
the equipotential bonding conduc-
tor using another equipotential
bonding cable with a gauge of at
least 4 mm
²
.
t Connect the cable lug of the
equipotential bonding cable to
the grounding terminal of the
ex-link converter.
t Connect the equipotential
bonding cable to the custom-
er-supplied equipotential
bonding conductor.
3.6 Establishing the Power Supply
Required qualification: Electrician
The scale is connected to the power supply via a PC/
notebook or using the optional AC adapter YPS06-USB
(see Chapter “6 Accessories”, page 10), which is supplied with
mains adapters for use in various countries.
NOTICE
Ensure that the voltage rating printed on the AC adapter
is identical to your local mains voltage (connection data
see Chapter “8.1 General Data,” page 11).
If the stated supply voltage or the plug design of the AC
adapter does not comply with your country's standard,
please inform your nearest Sartorius representative.
Power supply via the AC adapter is only required:
When no PC or notebook is available.
In exceptional cases, when the output voltage of the
USB interface of the PC or notebook is not sufficient.
AC adapter assembly is described in the following.
3.6.1 Installing AC Adapter
t
WARNING Lethal electric shock and equipment
damage due to incorrect power plug adapter! Only use
the country-specific power plug adapter. Never plug
the power plug adapter into the socket when it is
disconnected from the AC adapter.
t Use the right mains adapter for your mains power supply:
Mains adapter set YAK01
Bag Region/Country
a) transparent Europe/EU (except United Kingdom)
b) blue USA
c) yellow United Kingdom
Mains adapter set YAK02
d) red Australia
e) turquoise South Africa
f) white Argentina
g) pink Brazil
Mains adapter set YAK03
h) light brown China
i) black India
j) green Korea
t Push (1) and slide (2) the mains
adapter required for your power
supply into the opening of the
AC adapter module.
When doing this, the mains
adapter needs to lock into
position.
Removing/Replacing the Mains
Adapter
t Unlock (1) and then remove (2)
the mains adapter.
EVO1X | LAB1X Installation Instructions 7