1
Connecting a PC/Notebook
t Insert a USB cable (1) into
the right-hand socket on the
ex-link converter and connect
the cable to a PC or notebook.
If the ex-link converter
YCO14-Y is being used and
installed in Zone 2 or Division 2
potentially explosive areas and
the USB cable (1) does
not have an intrinsically safe
electrical circuit, the cable
must be secured against
disconnection. Refer to the
Control Drawing 2021459 and
Safety Instructions 2023040
from Page 53 as well as the
following illustrations.
1
2
2
t Fasten the
strain relief
YSR01 to the
converter (1)
using the two
screws (2).
2
1
2
3
t Attach the
USB cable to
the strain
relief (1)
using the
clamp (3)
and the two
screws (4).
6
5
t Use the
protective
cap (6) to
seal up the
left-hand
USB port (5).
2
Connecting the AC Adapter
(Optional); not for YCO14-Y
when Installed in Zone 2 or
Division 2 Potentially Explosive
Areas
t Insert an additional USB cable
(2) into the ex-link converter.
t Insert the USB cable into the
YPS06-USB AC adapter.
t Plug the AC adapter into a wall
outlet (supply voltage).
3.7 Anti-theft Locking Device
t If required, secure the scale at
the back.
3.8 Leveling
Leveling for PMA.HD and Verified Models
Leveling the scale compensates for slant or unevenness at the
place of installation. The scale must be perfectly horizontal to
ensure consistent, reproducible weighing results.
The scale needs to be re-leveled and then adjusted if
necessary each time its setup location is changed.
12*
9*
6*
* o'clock
t Turn the two leveling feet as
shown in the illustration until
the air bubble is centered within
the circle of the level indicator.
Air bubble at “12 o'clock:” Turn
both feet clockwise.
Air bubble at “3 o'clock:” Turn
the left foot clockwise and the
right foot counterclockwise.
Air bubble at “6 o'clock:” Turn
both feet counterclockwise.
Air bubble at “9 o'clock:” Turn
the left foot counterclockwise
and the right foot clockwise.
8 EVO1X | LAB1X Installation Instructions