Calibration/Adjustment
12
You can calibrate/adjust the scale by pressing the
U key (5).
To do so, menu code 1 5 1 must be selected.
Calibration weight: 5,000 g; accuracy: + 0.075 g.
After connection to AC power and before each
calibration/adjustment, allow the scale to warm up
for approx. 30 min.
Hold down the U key (5) for 2 sec. When 5000 is
displayed, release the key.
Center the calibration weight on the weighing pan
(11).
Calibration/adjustment is performed automatically.
After calibration and adjustment, remove the
weight.