SARTORIUS PMA 7500...
5
Vorein03.EPS
Vorein04.EPS
Operating the Scale
Function of Keys
e On/off. The scale remains in the standby mode.
= Tare / Zero key (display is set to 0,0 g)
w Toggle. Toggles the display between weight and
calculated value; also for changing Reference
sample quantity/reference percentage/ numbers
of subweighing operations.
v Funktion key (start application program)
c Clear function key, Used to interrupt/cancel functions, delete input.
Y Curser key up
y Curser key down
Recalculations key (start recalculation program)
Memory key (store values)
Scale Operating Menu
Accessing the Scale Operating Menu and Changing the Settings
Press e to turn the scale off and press again to turn it back on.
Press = briefly while all segments are lit (self-test).
–“
1.1.
1.1.
1.” is displayed.
Important: You can configure software settings to block access to the scale operating
menu (menu code: 8 1 2). If menu access is blocked, this is indicated by a
minus sign preceding the menu code number (e.g., ”
- 1.- 1.
- 1.- 1.
- 1.” rather than “
1.1.
1.1.
1.”).
To make the menu accessible, select the menu code 8 1 1 (the only menu
code that can be selected when menu access is blocked). This setting can be
changed again the next time you activate the scale operating menu.
Selecting a Menu Code
To select a certain menu code, press = to change the number of the
digit displayed to the desired code number. Numbers change in a
cyclical sequence, starting at 1 again after 9 is reached.
To select the next digit (menu level), press w
Once the first 2 digits are selected, the entire 3-digit code setting is
displayed, with a small superscript “o” indicating that this is the current
setting.
Press w to return to the higher menu level
Changing a Menu Code Setting
Once you have selected the desired code setting as described above,
press w for >2 seconds to confirm the setting.
The superscript “o” is now displayed to indicate that this is the current
setting.
Vorein01.EPS
Vorein02.EPS
Anz-rec.TIF
Anz-norm.TIF