6
Exiting the Scale Operating Menu with or without Storing the New Settings
To exit the scale operating menu and store the new settings:
Press = for > 2 seconds.
To exit the scale operating menu without storing any changes to the
settings: Press e.
Overview of the Scale Operating Menu Codes
See pages 12 for an overview of the scale operating menu codes.
Activating the BPI Mode
BPI = Binary Processor Interface (special Sartorius data transfer protocol)
SBI = Sartorius Balance Interface (RS-232 communication)
The scale must be in the BPI mode in order to use the CAS programs with the
MC-1 server (version 4.6 or later) or a personal computer (version 1.33 or
later). These programs are required especially for:
adjusting linearity and/or span when the adjustment error is > 2%,
or
replacing the main PCB.
Important: After working in the BPI mode, make sure to set the write-protect again
(using the “Close” function) so that the scale returns to the standard data
transfer protocol (SBI mode). Otherwise you will not be able to operate
peripheral devices with the scale; when p is pressed, the error code
E 30E 30
E 30E 30
E 30” is displayed.
Procedure:
Disconnect the equipment from the power supply
Open the unit (see page 9)
Re-connect the equipment to the power supply
Plug in the mode adapter (6740-63)
The black protective cap points toward the outer edge of the scale.
Note: An incorrect postion has no affect to the scale.
The scale now performs its self-test (display segment check) repeatedly.
Wait until the segment check has lit up at least 3 times; the scale is now in
BPI mode.
Remove the mode adapter
Close the scale housing (see page 9)
You can now use the scale with the service software.
Vorein01.EPS