EasyManua.ls Logo

Sartorius PMA7500-X - Replacing the Display PCB; Replacing the Strain-Gauge Load Cell; Load Cell Replacement Procedure

Sartorius PMA7500-X
16 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
Replacing the Display PCB
The display PCB must be replaced if:
the display is defective
data output is defective
When replacing the display PCB, replace the cable connecting the display
PCB to the main PCB at the same time.
Open the scale (see page 9)
Remove or cut the wire tie holding the cable that connects the electronics
module to the display and control unit
Now the display head together with the suport arm can be removed
Carefully remove the protective cap (A) using a proper tool
Push the pin out of the support arm
Now the defective display head together with the connecting cable can
be removed
Follow the above instructions in reverse order to install the new PCB
Replacing the Strain-Gauge Load Cell
Open the scale (see page 9)
Unplug the load-cell cable from the main PCB
Remove the two screws from the pan support and put it aside
Remove the three screws that tighten the strain gauge load cell from the
underneath. now the defective strain gauge load cell can be removed.
Place the new strain gauge load cell in the housing adjust its position and
tighten it again.
Tighten the pan support again.
Guide the connecting cable into the electronics module and plug the
cable into the main PCB
Now the zero-point offset value the lineaity and the span need to be
checked / adjusted.
Kopf-01d.TIF