Setup and Configuration  PR 5210 Operating Manual 
28  Sartorius 
4.8.3 Bus size 
The normal bus size is 8. 
10 byte bus size is used for a coded data transfer. 
Default is 8. 
 
 
4.8.4 Communication 
The serial line could be used for data communication with a remote display (e.g. PR1627 or PR1628) or 
a SMA data protocol.  
Default is 'off'. See chapter 2.5.3. 
 
 
4.8.5 Baudrate 
The baudrate can be selected between 300, 600 ..... and 19.200, care has to be taken that the same 
transfer speed is set at the remote display / terminal. Default is 9600. 
See chapter 2.5.3. 
 
 
4.8.6 Access 
The access code can consist of 9 decimal digits at maximum. If the access code is set to 0, no check on 
the code is done. If an access code has been set (has to be entered twice for security reason), data and 
parameters cannot be altered, without entering the code. As long as the user has got access, he is 
allowed to change the access code.  
Default is 0. 
 
 
4.8.7 Outputs 
The following selections are possible per each of the 3 digital outputs: 
transparent 
The PLC is controlling the output 
ADU error 
Output is set if ADU is in error state 
Limit1 
Result of comparison of limit 1 values with actual weight (see chapter 4.8.9) 
Limit2 
Result of comparison of limit 2 values with actual weight (see chapter 4.8.9) 
Limit3 
Result of comparison of limit 3 values with actual weight (see chapter 4.8.9) 
Tare active 
Output is set, if transmitter is in net mode 
 
 
4.8.8 Inputs 
The following selections are possible per each of the 3 digital inputs: 
none 
Input is not used for transmitter control 
set zero 
The transmitter will be set to zero (see chapter 4.7.2.10). 
(standstill has to be fulfilled, else standstilltimeout will be given, see chapter 4.7.2.5 , 4.7.2.7 
and 4.7.2.9) 
set tare 
The transmitter will be switched to net mode. 
(standstill has to be fulfilled, else standstilltimeout will be given, see chapter 4.7.2.5 , 4.7.2.7 
and 4.7.2.9)  
reset tare 
The transmitter will be switched to gross mode