PR 5210 Operating Manual  Setup and Configuration 
Sartorius  31 
4.10  Analog Output Adaption 
For PR 5210/00 und PR 5210/10 only. 
 
 
 
The analog output current is at the receiver side usually routed over a resistor, measured as voltage 
and the digitalisied. The errors of this transfer chain could be compensated at the transmitter side. 
With this menu the points for 4 and 20 mA could be individually adapted, so that they are precise 
again at the receiver side. 
 
The adaption is done in several steps: 
 
1. step:  Press [Start Adaption]. 
2. step:  The analog output is set to 4mA. Measure the current at the receiver side and enter the 
value of the measured current. E.g. 4.002mA.. Press [next]. 
3. step:  The analog output is set to 20mA. Measure the current at the receiver side and enter the 
value of the measured current. E.g. 20.003mA.. Press [next]. 
 
The current is now corrected to 4.000mA and 20.000mA.