PR 5210 Operating Manual  Profibus Interface 
Sartorius  49 
 
6.1.6.4  Reset transmitter to default (factory) data 
Master 
Transmitter  
Set parameter P20 to 1 (Start Cal) 
(method see chapter 6.1.6.1) 
The CAL switch is tested, if the switch is not in the 
'locked' position, the procedure is started. 
Set parameter P20 to 2 (default = factory 
data) 
(method see chapter 6.1.6.1) 
Default data are taken as calibration data. 
All ADU data will be reset. See chapter 4.7 
Set parameter P20 to 3 (SaveAndExit) 
(method see chapter 6.1.6.1) 
Data are stored in the non-volatile EAROM