Below-Cell Weighing
A port for a below-cell weighing hanger is located on the bottom of the weigh cell
(not on model WZA25-NC).
§ Remove the closing screw.
$ Overload protection
Standard feature on the following models: WZA224-N/-NC, WZA1203-N/-NC,
WZA523-N/-NC
! No overload protection provided on the following models:
WZA8202-N/-NC, WZA25-NC
§ Carefully install the customer-specific hook.
Threaded fastener for hook: M3
Maximum torque: 0.8 Nm.
! Maximum screw installation depth: do not exceed 5 mm!
$ Install a draft shield if necessary