EasyManua.ls Logo

Sartorius WZA N Series - Parameter Settings: Overview

Sartorius WZA N Series
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
Parameter Settings: Overview
ο = Factory setting; = User-defined setting
Level 1 Level 2 Level 3 Level 4 Code
1.) Setup Bal.Scal Ambient V.Stable Very stable conditions 1. 1. 1. 1
Weigh cell conditions ο Stable Stable conditions 1. 1. 1. 2
functions (Filter adaptation) Unstabl Stable conditions 1. 1. 1. 3
V.Unstbl. Very unstable conditions 1. 1. 1. 4
App.Filt. ο Final.Rd. Final readout mode 1. 1. 2. 1
Application filter Filling Filling mode 1. 1. 2. 2
Stab.Rng. 1/4 dig. (digit) 1. 1. 3. 1
Stability range 1/2 dig. (digit) 1. 1. 3. 2
1 dig. (digit) 1. 1. 3. 3
ο  2-dig. (digit) 1. 1. 3. 4
4 dig. (digit) 1. 1. 3. 5
8 dig. (digit) 1. 1. 3. 6
Stab. delay No delay 1. 1. 4. 1
ο Short delay 1. 1. 4. 2
Medium delay 1. 1. 4. 3
Long delay 1. 1. 4. 4
Taring W/o Stb W/o stability 1. 1. 5. 1
Taring ο  W/ StAb After stability 1. 1. 5. 2
Aut.zero off 1. 1. 6. 1
Auto zero ο  on 1. 1. 6. 2
Wt.Unit For list of units, 1. 1. 7. 1
Basic Gram to Newton to
Weight unit 1. 1. 7.23
Display ο  all 1. 1. 8. 1
Basic accuracy Minus 1 One level lower 1. 1. 8. 2
Increment of the measured values one level lower 1. 1. 8. 3
Increment of the measured values two levels lower 1. 1. 8. 4
Increment of the measured values three levels lower 1. 1. 8. 5
Incrm. 1 Last digit single increment 1. 1. 8. 6
resolution by a factor of 10 1. 1. 8. 8
Cal./Adj. ο cal. ext. Adjustment/calibration with factory-set weight 1. 1. 9. 1
Function of the ο e.cal.usr. External calibration/adjustment
Q key with user-defined weights (factory-set on WZA25-NC) 1. 1. 9. 3
Cal.Int. Internal calibration/adjustment only on models WZA…-NC 1. 1. 9. 4
lin.ext. Linearization with factory-set weights 1. 1. 9. 6
lin.e.usr. Linearization with user-def. weights 1. 1. 9. 7
str.prel. Set preload 1. 1. 9. 8
clr.preload Clear preload 1. 1. 9. 9
blocked Q Blocked 1. 1. 9. 11
Cal.routine ο Adj. one sequence 1. 1. 10. 1
Cal./adj. as required 1. 1. 10. 2
Cal.Unit ο Grams 1. 1. 11. 1
Weight unit Kilogr. Kilograms 1. 1. 11. 2
for calibration Pounds 1. 1. 11. 3
Zero rng. default. (factory-set) 1. 1. 12. 1
Zero range ο 2 Perc.ent 1. 1. 12. 2
5 Perc.ent 1. 1. 12. 3
10 Perc.ent 1. 1. 12. 4
Int.zero Zero at power-on default (factory-set) 1. 1. 13. 1
Power On Initial zero 2%/max. cap 1. 1. 13. 2
ο Initial zero 5%/max. cap 1. 1. 13. 3
Initial zero 10%/max. cap 1. 1. 13. 4
Initial zero 20%/max. cap 1. 1. 13. 5
Initial zero 50%/max. cap 1. 1. 13. 6
Initial zero 100%/max. cap 1. 1. 13. 7
on.tare ο On 1. 1. 14. 1
(Tare/Zero at off 1. 1. 14. 2
Power/Zero-setting range)
Output rate ο Normal 1. 1. 15. 1
Fast (five times faster) 1. 1. 15. 2

Related product manuals