Leveling the Weigh Cell in a Portable Weighing System (Leveling Feet Optional)
Purpose:
To compensate for uneven areas at the place of installation.
To ensure that the weigh cell is placed in a perfectly horizontal position for consistently
reproducible weighing results.
Always level the weigh cell again any time after it has been moved to a different location.
§ Adjust the leveling feet until the air bubble is centered within the circle on the level
indicator.
Permanently Installed Weigh Cells
Adjust the weigh cell after it has been installed in the system in its permanent
location – see next page.
The weigh cell must be calibrated/adjusted again any time its location or position
is changed.
For optimum operation, install the weigh cell in a horizontal position.
1) Bottom plate of the weigh cell
2) Fastening frame of the system
§ Fastening with M6 screws:
Connection to the threaded fasteners on the weigh cell (1): torque 2.5 Nm
§ Fastening with M4 screws:
Connection to the threaded fasteners of a user-specific frame (2).
! Do not unscrew the sleeve.