Parameter Settings: Overview
ο = Factory setting; = User-defined setting
Level 1 Level 2 Level 3 Level 4 Code
1.) Setup Bal.Scal Ambient V.Stable Very stable conditions 1. 1. 1. 1
Weigh cell conditions ο Stable Stable conditions 1. 1. 1. 2
functions (Filter adaptation) Unstabl Stable conditions 1. 1. 1. 3
V.Unstbl. Very unstable conditions 1. 1. 1. 4
App.Filt. ο Final.Rd. Final readout mode 1. 1. 2. 1
Application filter Filling Filling mode 1. 1. 2. 2
Stab.Rng. 1/4 dig. (digit) 1. 1. 3. 1
Stability range 1/2 dig. (digit) 1. 1. 3. 2
1 dig. (digit) 1. 1. 3. 3
ο  2-dig. (digit) 1. 1. 3. 4
4 dig. (digit) 1. 1. 3. 5
8 dig. (digit) 1. 1. 3. 6
Stab. delay No delay 1. 1. 4. 1
ο Short delay 1. 1. 4. 2
Medium delay 1. 1. 4. 3
Long delay 1. 1. 4. 4
Taring W/o Stb W/o stability 1. 1. 5. 1
Taring ο  W/ StAb After stability 1. 1. 5. 2
Aut.zero off 1. 1. 6. 1
Auto zero ο  on 1. 1. 6. 2
Wt.Unit For list of units, 1. 1. 7. 1
Basic Gram to Newton to
Weight unit 1. 1. 7.23
Display ο  all 1. 1. 8. 1
Basic accuracy Minus 1 One level lower 1. 1. 8. 2
Increment of the measured values one level lower 1. 1. 8. 3
Increment of the measured values two levels lower 1. 1. 8. 4
Increment of the measured values three levels lower 1. 1. 8. 5
Incrm. 1 Last digit single increment 1. 1. 8. 6
resolution by a factor of 10 1. 1. 8. 8
Cal./Adj. ο cal. ext. Adjustment/calibration with factory-set weight 1. 1. 9. 1
Function of the ο e.cal.usr. External calibration/adjustment
Q key with user-defined weights (factory-set on WZA25-NC) 1. 1. 9. 3
Cal.Int. Internal calibration/adjustment only on models WZA…-NC 1. 1. 9. 4
lin.ext. Linearization with factory-set weights 1. 1. 9. 6
lin.e.usr. Linearization with user-def. weights 1. 1. 9. 7
str.prel. Set preload 1. 1. 9. 8
clr.preload Clear preload 1. 1. 9. 9
blocked Q Blocked 1. 1. 9. 11
Cal.routine ο Adj. one sequence 1. 1. 10. 1
Cal./adj. as required 1. 1. 10. 2
Cal.Unit ο Grams 1. 1. 11. 1
Weight unit Kilogr. Kilograms 1. 1. 11. 2
for calibration Pounds 1. 1. 11. 3
Zero rng. default. (factory-set) 1. 1. 12. 1
Zero range ο 2 Perc.ent 1. 1. 12. 2
5 Perc.ent 1. 1. 12. 3
10 Perc.ent 1. 1. 12. 4
Int.zero Zero at power-on default (factory-set) 1. 1. 13. 1
Power On Initial zero 2%/max. cap 1. 1. 13. 2
ο Initial zero 5%/max. cap 1. 1. 13. 3
Initial zero 10%/max. cap 1. 1. 13. 4
Initial zero 20%/max. cap 1. 1. 13. 5
Initial zero 50%/max. cap 1. 1. 13. 6
Initial zero 100%/max. cap 1. 1. 13. 7
on.tare ο On 1. 1. 14. 1
(Tare/Zero at off 1. 1. 14. 2
Power/Zero-setting range)
Output rate ο Normal 1. 1. 15. 1
Fast (five times faster) 1. 1. 15. 2