FAULT POSSIBLE CAUSE CORRECTIVE ACTION
Nothing working;
the green
indicator lamp
doesn't light up
The ON/OFF switch is on O Turn the ON/OFF switch to I
Mains lead not properly connected Check the mains plug
No electric current Check electric current supply
Fuse(s) not working
Change the external fuses. Caution: the unit includes a
third, internal, fuse that is inaccessible to the user.
Contact the dealer
The green
indicator lamp
lights up but the
orange indicator
lamp does not,
and the buzzer
does not work
Footswitch lead not properly
connected
Check the footswitch: plug in firmly
Faulty footswitch Replace the footswitch or contact the dealer
Thermal protection of device Wait for the device to cool down
The indicator
lamps come on
and the buzzer is
working, but
there is no high-
frequency
current
Electrode-holder not properly
connected
Check the electrode-holder cord:
- Push unit-side connector in properly
- Push electrode-holder-side connector in properly
Capacitive coupler not properly
connected
Check the capacitive coupler: the connector must be all
the way in
Other possibilities Contact the dealer
The electrode
incises with
difficulty or is
not incising at
all
Capacitive coupler in the wrong
position.
Follow the indication Patient Side printed on the
capacitive coupler.
Make sure the entire surface of the capacitive coupler is
in contact with the lightly- clothed patient, for example,
on the scapula.
Soiled electrode Set the unit to O. Clean the electrode.
Electrode moving too fast Reduce operating speed
Unsuitable electrode
Select the appropriate electrode for the operating
procedure
Electrode worn Replace the electrode
The electrode
sticks to the
tissue
Power set too low
Increase power up to the incision threshold. There is no
need to go beyond this threshold
Inappropriate electrode
Select the appropriate electrode for the operating
procedure
The electrode
incises but there
are sparks
Incision power set too high
Reduce incision power down to the threshold. There is no
need to go beyond this threshold