EasyManua.ls Logo

sauter FC - Troubleshooting

sauter FC
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FC-BA-e-1410 15
Enter Calibration interface, as shown in Fig 5-1.
Calibration interface is shown in
Fig 5-2
Load a standard force. Now the value in standard input area is just equal to the
current measured value. Wait a moment for the force stabilization.
Press and to input the standard force value.
Press to enter the next calibration. Pressing can interrupt the calibration.
When the 3 times calibration had been finished or been interrupted, a confirm win-
dow will pop up for asking you "Save and Exit" (YES) or “Save and Exit” (NO), (Fig 5-
3a)
Press or to select, then press .
If "YES" is selected, "Calibrate complete!" is displayed, see 5-3(b).
5.4 Troubleshooting
If the force gauge in use processes such as errors, please accord to the following
way for troubleshooting, see table 5-3. Do not disassemble the gauge by yourself to
repair. If you cannot resolve the fault yourself, please contact our company.
5-3(a) 5-3(b)
Fig. 5-3
(a) (b)
(c) (d)
Calibration times
Current measuring value
Standard value input
Fig 5-2

Related product manuals