6
ADJUSTMENT
Nota : The manufacter parameters of the IPE50 unit are : 2mV/V -> 10.000 per 0.001
Attention : the default calibration parameters will probably not match with your application requirements. You
have to calibrate the system according to your needs.
If the unit is in trade-approved mode, you must open the unit to insert the jumper (see jumper CAL page 11 or
13)
To enter the SET-UP environment of the scale, switch on the unit and press the ->T<- key for an instant during
the countdown.
To leave the setup environment
, press the C key various times until the indicator shows “SAVE? on the
display, confirm with
↵ to memorize and return to weighing.
->0<- Decreases the selected digit (blinking) or go down.
->T<- Increases the selected digit (blinking) or go up.
MODE Selects the digit to be modified (blinking), from left to right.
C Quickly zeros the displayed value or comes back on the step before.
Valid the new value by pressing the key
↵.
1) Enter the SET-UP environment of the scale (when turned on, press the ->T<- key for an instant during
the countdown) .
2) TypE is displayed.
Press ↵, ind.Ch. is displayed, press ↵
Ind.Ch. Standard using for 1 to 4 independent channels , for DEP.Ch. Using for the connection of 2
to 4 identical sensors on a same system and TrAnSM. see user manual NU-IPE50-1-E-XX
3) Select SEtuP with key ->T<- then ↵, → ConFiG confirm with key ↵ → select CALib with key ->T<-
and press ↵.
4) NChAn is displayed, press ↵ -> Ch x where x is the channel number which must be adjusted.
(for 1 channel or 1 sensor, choose Ch 1 because the sensor must be connected in channel 1 (CELL1))
5) ParAM is shown, choose CALib with key ->0<- and press ↵.
6) “dECI” step (decimal point) press ↵
The selectable values are 1.0 (a decimal), 1.00 (two decimals), 1.000 (three decimals), 100000 (no
decimal); confirm with ↵
(!) 1.0
7) “u.M.” step press ↵
Set the unit of measure (g, Lb, t, kg) and press
↵
(!) kg
Calibration with load on the system, for theorical calibration (with loadcell parameters, less accuracy)
see paragraph 18)
8) “diU” step press ↵
Set the minimum division and press ↵ (selectable values: 1, 2, 5, 10, 20, 50).
Ex: value 1 : the right digit could change from 1 to 1 : 0 ; 1 ; 2 ; 3 ; 4 ; 5 ; 6 ; 7 ; 8 ; 9
value 2 : the right digit could change from 2 to 2 : 0 ; 2 ; 4 ; 6 ; 8
value 5 : the right digit could change between 0 and 5 only
value 10 : the right digit will be always 0 and the second one from the right could change from 1 to