21) DEAd.ld (dead load)
If you want to enter the value of the system dead load (empty silo weight for ex) press in other case
press ->0<- (kno.WGt is displayed see paragraph after)
You could enter the weight dead load value with the keys
If no number after the coma has been selected, you could set value from -99999.9 to 99999.9
If 1number after the coma has been selected, you could set value from -9999.99 to 9999.99
If 2 numbers after the coma has been selected, you could set value from -999.999 to 999.999
If 3 numbers after the coma has been selected, you could set value from -99.9999 to 99.9999
Press
Or
22) Kno.WGt (enter the value of the know weight on the system) in another case press the ->0<-
(CEL.Sen is displayed, press key C, th.CAL? is displayed, press to memorize, rAnGE1 is displayed,
see paragraph 23).
If the system is loaded with a know weight, you can set the unit with this weight (with the keys). The
unit will memorize the loadcells signal and the weight Press , GET.Wt? is displayed, press .
Wait is displayed for several seconds (the unit memorize the loadcells signal) then Kno.WGt which is
blinking and after the unit will go to the step CEN.Sen, press C, th.CAL? is displayed, press ,
rAnGE1 is displayed.
23) Press the C key various times until the indicator shows “SAVE? in the display:
Confirm with to memorize and return to weighing.
QUICK CALIBRATION OF ZERO
It is useful to calibrate just the point of ZERO when a permanent tare weight is put onto the platform (for
example a roller unit).
1) Enter in the SET-UP environment of the scale (when turned on, press for an instant the TARE key).
2) Enter in the SEtuP ConFiG 0.CALib step and press key (the display shows “CAL.0?”).
3) Put the tare on the scale and press key to confirm the operation.
4) Once the zero calibration is made, press many times the C key until the indicator shows “SAVE? in the
display: confirm with
to store and return to weighing.