ERROR MESSAGES
While using the indicator, it is possible to incur in the following errors:
MESSAGE DESCRIPTION
ADC Err
Analog/digital convertor doesn't work. Check the wiring of the loadcells. Measure the signal
output of the loadcells between the pins + and - sig, the tension must not exceed 15mV.
PREC.
It is displayed if one tries to calibrate a point without first having confirmed the number of
calibration points
ERMOT Weight unstable during the acquisition of a point during calibration.
ERPNT During the acquisition of a calibration point a null value has been read by the converter.
Er – 11
Calibration error: a too small sample weight has been used; it is advisable to use a weight
equal to at least half of the scale capacity.
Er – 12
Calibration error: the acquired calibration point (tP1 or tP2 or tP3) is equal to the zero point
(tP0).
Er – 37
The converter points are less than the instrument's internal divisions.
The sensor signal is negative
The load has not been applied on the step TP1
Er – 85
It is displayed when the instrument has not yet been calibrated and initialized.
Press the ->T<- key when the instrument displays “ERR – 85” to enter the technical set-up
environment (with the standard procedure it is not possible) programming of all the
parameters of the set-up environment and the calibration.
C.Er.-36
During the calibration some internal negative points have been calculated:
- the calibration point is less than the zero point
- the signal is negative (check the connections)
C.Er.-37
During the calibration some internal points less than the minimum value have been
calculated:
- The calibration point is equal to the zero point
- A capacity too high in relation to the division has been set
HW-Err
Hardware error. Software not compatible with the installed hardware.
The hardware expansion is missing which allows the software to function.
ELECTRICAL CONNECTION
Nota for IPE50 DIN and XLI:
With the connector CELL1, it is possible to connect a sensor with 4 or 6 wires (with regulation of the sensor
power supply).
With the connectors CELL 2 to 4, the regulation of the sensor power supply can't work (there is only 4
connection)
You must connect at least one sensor on CELL1.