Operating Instruction Procedure in case of an alarm 3.7
Art. no.: 2.510474 rev.: d
ARGUS LCM/PLUS
3.7.3 Suppressing/acknowledging an audible alarm
There are two ways to proceed in the case of an alarm. The procedure depends on
the set alarm stop mode (On/Off) and the alarm suppression time in the menu
About+/Alarm:
(1) Alarm stop off (suppression)
This function suppresses an audible alarm for a defined period of time. However, the
flashing measurement field (red measured value and coloured field) will remain.The
audible alarm is reactivated after the defined period of time has elapsed.
(2) Alarm stop on (acknowledging)
This function will suppress an audible alarm as long as the defined limits are exceed-
ed. However, the flashing measurement field (red measured value and coloured field)
remains.