Page 73
Technical Data 8
User Guide ARGUS LifePoint sensor 8.6
Art. no.: 2.511279 Rev. d
FRED
®
easyport
®
plus
8.6 ARGUS LifePoint sensor
Sensor name
Argus LifePoint
Dimensions
80 x 25 mm (diameter/ depth)
Weight
152 g
Cable length
2 m
Power supply 5 VDC via USB from the medical device
Environmental conditions
Operation
Transport / Storage
For Transport and storage
between uses
Note: The environmental conditions for the LifePoint depend on the device and are
determined by it)
• -5 ºC ... 50 ºC relative humidity at 10 - 95 % (non condensing)
• -10...50°C / +5...50°C, humidity 10...95% (non condensing), pressure
500...1060 hPa.
• -40... 75 ºC, relative humidity of 10...95 % (non-condensing)
Atmospheric pressure 500...1060 hPa (5000m to -400m) for above
Measured values
Frequency
Compression depth
1 to 160 compression/min (cpm)
1 to 127 mm
Accuracy
± 3 compressions/min
± 5% at 50 mm (laboratory condition)
Life cycle
500’000 compressions
Dust and water protection IP66
Protection class Type BF defibrillation proof