4.7. Options
The settings, which will be valid for the entire treatment
can be selected in the menu O pt io ns .
The setting of the curvature of the root canal serves
for the calculation of the wear.
The following can be selected as r o ot - c u r v a t u r e :
easily (Factor 1x)
medium (Factor 2x) – pre-set
strong (Factor 3x).
A pe x - f u n ct io ns , here you can select (see chapter 4.7.1):
auto-stop (arbitrary, stopping time: 0,5, 1, or 2 seconds)
no auto-stop
off
audible assistance, here you can select:
off
closer to apex - the closer the file gets to the Apex,
the shorter the time between the tones.
Motor Load - the stronger the torque-load (file is approaching the
maximum torque limit), the shorter the times between the tones.
The function 1:1 calibrate contra-angle so should be used after each change and each sterilization of the
contra-angle. This can compensate smaller torque-losses at the contra-angle and helps to ensure a stable usage of
small torque-limits. If the calibration is not possible, the contra-angle is dirty or damaged. In this case, please contact the
manufacturer.