EndoPilot
*related to the additional module “CoPilot” - 32 - 609 2201 v12 www.schlumbohm.de
Part 2 – Continued …
Cleaning: manual:
B. (all parts for steam sterilization, except file cable see above)
1. lip clip
2. nut (screw cap of the Down-pack hand piece), O-ring (blue) as well as pen tip
(heating instruments)
3. heat insulator and needle (the disposable needles are to be processed before
usage) (see chapter “Needle”)
-Material: Combined cleaning and disinfection agent ID213
1
disposable tissues, ultrasonic
bath with instrument-tank, brush for interior drillings 3 mm (no metal bristles)
-Process: for initial disinfection submerge all parts intended for processing in a fresh bath
with combined cleaning and disinfection agent ID213
1
. This procedure is necessary for
protecting your safety and it stops the drying of adhesions. Please adhere to the
concentrations indicated by the manufacturer. Do not submerge the parts overnight, do not
keep the parts longer than 4 hours in the pre-cleaning-bath.
-For removing contaminations from the components use disposable tissues and cold
running water.
NOTE: Use an appropriate brush when cleaning drills. Wear protective clothing.
-Process: submerge the parts in a fresh solution with combined cleaning and disinfection
agent ID213
1
in the ultrasonic bath. The parts have to be entirely covered and may not
touch each other. Please adhere to the concentrations indicated by the manufacturer.- To
entirely clean the components from contaminations, use a soft brush and clean all inner as
well as outer surfaces several times. Examine the parts to ensure appropriate cleansing.
The cleaned parts have to remain in the bath for the exposure time indicated by the
manufacturer.
Needle (new) take one needle out of the transport packing, wear clean disposable gloves.
Place the needle separately in a clean, disinfected vessel with fresh combined cleaning
and disinfection agent ID213
1
in the ultrasonic device. Adhere to concentrations and
exposure times indicated by the manufacturer.
Cleansing and disinfection has to be carried out separately from the other parts to avoid
contaminating the needle with suspended solids.
-Rinsing: Take out all parts and rinse them thoroughly with de-ionized water containing low
levels of germs and endotoxines (aqua purificata). Use the ultrasonic bath, change the
water several times, it is vital to ensure that the cleaning and disinfection agents are
removed completely (for further instructions see C.).
Drying:
C. Wipe handpieces (or gun) and cables with a disposable tissue. In addition blow dry the
file cable with file clip, lip clip, nut (screw cap), O-ring and pen-tip (heating instrument), as
well as needle, heat-cover, guide-cylinder and the piston of the BackFill gun with oil-free,
dry and germ-low compressed air (use a sterile filter).
Maintenance: No maintenance necessary
Insert the piston with the black knob and the guide-cylinder of the B-Fill gun into the gun.
To do so press the locking knob and insert the guide cylinder, the nose pointing upwards,
into the gun, as far as possible. Turn now the knurled knob 90° to the right. Push the
knurled knob entirely into the housing. Turn now the knurl knob for another 180° to the
right.
Inspection and checking of
the function:
All parts: Inspect the parts for signs of damage and wear.
Do not use damaged components.
Packing: All parts intended for steam sterilization: separately in standardized sterile-bags for steam
sterilization. Packaging has to be sufficient.
The disinfected handpieces (or gun) and Apex cable (main-cable with plug) can be stored
in a plastic bag (unsealed / no autoclaving)
Attention ! Do not autoclave other parts than the intended.
Sterilisation: Steam sterilization (fractioned vacuum) autoclave device class B: 5 min at 134°C, 2 bar.
Treatment duration should be sufficient for the potential germs, if necessary, choose the
prion-program (for example: Melag Vakuklave 43 B)
Storage: No special requirements.
Additional information: When sterilising several instruments in one cycle, maximum load capacity of the autoclave
must not be exceeded.
Contact to manufacturer: For questions please contact your dealer or the manufacturer www.schlumbohm.de
The instructions stated above have been validated as SUITABLE by the medical device manufacturer for the preparation
of a medical product prior to its re-usage. The person who does the treatment is responsible that the treatment that
actually took place with the equipment, material and personnel used in the treatment institution leads to the desired
results. To ensure this it is normally necessary to validate and routinely supervise the process. Each deviation from the
instructions provided should be evaluated carefully by the person executing the treatment with regard to effects and
negative consequences.
1
cleaning and disinfection agent ID213 (www.duerrdental.com)