EasyManua.ls Logo

Schmidt EAR Series - Error Messages; Calibration of the Instrument

Schmidt EAR Series
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
3.5 Calibration of the Instrument
The balance is supplied with a 200 g calibration weight for recalibrating the instrument on
site or checking the measurement accuracy.
Requirement:
Instrument switched off as described in chapter 3.2.2.
The calibration weight is adjusted according chapter 3.3.1
There is no sample or container on the platform.
To calibrate the instrument:
- Press the button.
The display shows .
- Press the button for 3 seconds
- The display shows momentarily the adjusted calibration weight and
than is flashing.
- Place the calibration weight on the platform.
- Wait until the display has stabilized and shows .
- Remove the calibration weight from the platform.
After the calibration has occured the display shows momentarily and
than .
If the calibration occurs not with the adjusted calibration weight, the display shows
momentarily and than .
3.4.3 Error Messages
Overload:
The display shows .
The weight on the platform exceeds the measuring range.
Underload:
The display shows .
The platform has not been placed on the instrument.
Remedy:
- Place the platform on the support plate.
Zero setting not possible:
The switch-on procedure of the instrument is interrupted.
The display shows .
The platform is not placed on the balance or a weight greater than 20 % of the
measuring range has been placed on the platform.