EasyManua.ls Logo

Schmidt EAR Series - Function Settings; Zero Setting Options

Schmidt EAR Series
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
3.3 Settings
General information:
The settings for:
- Zero setting (auto zero)
- Timeout
- Display backlight
- Calibration weight
of the balance can be changed in the following way.
Requirement:
Instrument switched off as described in Chapter 3.2.2.
Zero setting:
- Press and hold the button and switch on the instrument with the button.
The display shows .
- By pressing the button, you can now choose between
Manual zero setting
Automatic zero setting
No function .
The default setting is automatic zero setting.
- You can now either return to the measuring mode by pressing the button
five times
or
Press the button once to set the
Timeout:
- By pressing the button, you can now choose between
No power off
Power off after 5 min.
Power off after 10 min.
Power off after 20 min.
Power off after 30 min.
of non-use.
The default setting is power off after 20 min.
- You can now either return to the measuring mode by pressing the button
four times
or
Press the button once to set the
3.3.1 Function Settings