EasyManua.ls Logo

SCHOTT CG 853P - Page 33

SCHOTT CG 853P
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
109
Calibration interval
(Int 3)
The flashing sensor symbol reminds you to calibrate regu-
larly. After the selected calibration interval (Int 3) expires,
the sensor symbol flashes. Measurements can continue.
Note
To ensure the high measuring precision of the measuring
system, perform a calibration after the calibration interval
expires.
Setting the
calibration interval
The calibration interval (Int 3) is set to 180 days in the fac-
tory. The interval can be changed (1 ... 999 days):
1 Switch off the instrument.
2 Press <
ì> and hold down the key.
3 Press the <on/off> key.
The
display test
appears briefly on the display.
The instrument then switches automatically to the
configuration level.
4 Press the <run/enter> key until Int 3 appears on the
display.
5 Press <> <> to set the required time interval until
the next calibration.
6 Confirm with <run/enter>.
7 Change to the measuring mode by pressing <ì>.

Related product manuals