Operation
110
Setting up the cell
constant manually
To set up the cell constant manually, proceed as follows:
1 Press the <CAL> key repeatedly until LF CELL ap-
pears on the display.
2 Press the <run/enter> key.
3 Press the <CAL> key repeatedly until the cell con-
stant to be set, e. g. 0.475 cm
-1
is displayed:
4 Set up the cell constant to be used by pressing
<
><>, e. g. 0.614 cm
-1
.
5 To return to the measuring mode: press the <ì> key.