EasyManua.ls Logo

Scientifica MMBP - Page 29

Scientifica MMBP
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
|25
panel). This may lead to expansion of the headstage and
subsequent drift.
You should allow half an hour after switching off before
re-testing stability
Recording chamber: are there any cables, wires or tubing
connected to the chamber?
Such cables could implement an external force causing
movement of the chamber.
When using brain slices, ensure that the harp use to hold
down the slice does so efficiently.
Check for air currents around the system ?
Open windows or air conditioning can give rise to air
cu
rrents.
Cover the cage with a sheet or plastic bag to eliminate
any air currents then test for drift.
If you are still experiencing drift difficulties after following these
steps, please contact us for further assistance.

Related product manuals