436-GC/456-GC
510 Error codes
Page: 62
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Code Error Description Error level Cause Condition(s) GC action Action needed
49
Rear FID Flame Reigniting
PFPD Board Failure
50 Front PFPD Board Failure Alert Defect PFDP board
Electronics off (all)
H2 flow set to 0.0 mL/min
Replace PFPD board
51 Middle PFPD Board Failure Alert Defect PFDP board
Electronics off (all)
H2 flow set to 0.0 mL/min
Replace PFPD board
52 Back PFPD Board Failure Alert Defect PFDP board
Electronics off (all)
H2 flow set to 0.0 mL/min
Replace PFPD board
53
Becoming Not Ready (during run)
Event
-
-
-
54
Becoming Ready (during run)
Event
-
-
-
IOC Failure
55
IOC Bad Addre Chan
Event
-
-
-
56
IOC Bad Cmd
Event
-
-
-
57
IOC Stepper in Motion
Event
-
-
-
58
IOC Bad Parameter
Event
-
-
-
59
IOC Run mode error
Event
-
-
-
60
IOC no Detector
Event
-
-
-
61
IOC Command not Allowed
Event
-
-
-
62 IOC_EFC_BUFFER_OVERFLOW, Unit rebooted Event
Caused by too many requests requested by IOC to
EFC Interface print.
GC reboot the interface print
firmware
-
63 IOC_EFC_COMMUNICATION_ERROR, Unit Rebooted Event
Communication between EFC Interface Print and
EFC broken.
GC reboot the interface print
firmware
-
Zone Open Sensor (zone)
67 Column Oven Open Sensor Error Column Oven sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
68 Zone 1 Open Sensor Error Zone 1 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
69 Zone 2 Open Sensor Error Zone 2 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
70 Zone 3 Open Sensor Error Zone 3 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
71 Zone 4 Open Sensor Error Zone 4 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
72 Zone 5 Open Sensor Error Zone 5 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
73 Zone 6 Open Sensor Error Zone 6 sensor not in use, mis used
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Shorted Sensor (zone)
74 Column Oven Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
75 Zone 1 Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
76 Zone 2 Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
77 Zone 3 Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
78 Zone 4 Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
79
Zone 5 Shorted Sensor
Error
Shorted sensor. Defect Sensor
Electronics off (all)
Check zone (5)